The latest food, drug and medical device recalls reported by the U.S. Food and Drug Administration, newest first. Data updated Oct 21, 2026.
87,918 recalls · page 2 of 30
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Reported Oct 21, 2026 · Z-0174-2027
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection
Reported Oct 21, 2026 · Z-0137-2027
Due to potential contaiminated of spefic lots of pneumonia and joint infection panels that could result in false positive results
Reported Oct 21, 2026 · Z-0150-2027
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Reported Oct 21, 2026 · Z-0001-2027
Following a Chromium-based web-browser, report and displayed images may not match.
Reported Oct 21, 2026 · Z-0186-2027
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Reported Oct 21, 2026 · Z-0165-2027
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Reported Oct 21, 2026 · Z-0175-2027
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Reported Oct 21, 2026 · Z-0185-2027
In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.
Reported Oct 21, 2026 · Z-0190-2027
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
Reported Oct 21, 2026 · Z-0141-2027
The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured.
Reported Oct 21, 2026 · Z-0146-2027
Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.
Reported Oct 21, 2026 · Z-0157-2027
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Reported Oct 21, 2026 · Z-0179-2027
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
Reported Oct 21, 2026 · Z-0161-2027
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
Reported Oct 21, 2026 · Z-0136-2027
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Reported Oct 21, 2026 · Z-0154-2027
Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
Reported Oct 21, 2026 · Z-0159-2027
The potential for the devices to be mislabeled with the incorrect length.
Reported Oct 21, 2026 · Z-0158-2027
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
Reported Oct 21, 2026 · Z-0135-2027
Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.
Reported Oct 21, 2026 · Z-0153-2027
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Reported Oct 21, 2026 · Z-0171-2027
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Reported Oct 21, 2026 · Z-0156-2027
Subrecall of a kitted device from SafeGuard Medical due to internal testing identified an issue affecting the device's sterile pad, potentially causing the sterile pad to degrade over time. The degradation may cause remnants of the nonwoven pad material to dislodge on the wound or in close proximity to the wound.
Reported Oct 21, 2026 · Z-0147-2027
Due to potential contaiminated of spefic lots of pneumonia and joint infection panels that could result in false positive results
Reported Oct 21, 2026 · Z-0149-2027
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Reported Oct 21, 2026 · Z-0169-2027
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Reported Oct 21, 2026 · Z-0182-2027
Affected kits were assembled with pediatric nasal cannula instead of the intended adult nasal cannula.
Reported Oct 21, 2026 · Z-0144-2027
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Reported Oct 21, 2026 · Z-0181-2027
Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.
Reported Oct 21, 2026 · Z-0139-2027
Labeling not compliant with 21 CFR 801 and steam sterilization IFU has been updated.
Reported Oct 21, 2026 · Z-0188-2027